Fracture of the tibial tray after total knee arthroplasty

dc.authorid0000-0003-1829-5700en_US
dc.contributor.authorAltıntaş, Faik
dc.contributor.authorŞener, Nadir
dc.contributor.authorÜğütmen, Ender
dc.date.accessioned2024-07-12T21:11:29Z
dc.date.available2024-07-12T21:11:29Z
dc.date.issued1999en_US
dc.departmentFakülteler, Tıp Fakültesien_US
dc.description.abstractWe report a case in which fracture of the metal tibial plate was thought to be caused by a combination of neutral anatomic rotation of the femoral component and an undersized tibial plate. This resulted in medial subsidence and baseplate fracture with loosening of the polyethylene and dislocation of the prosthesis.en_US
dc.identifier.citationAltıntaş, F., Şener, N. ve Üğütmen, E. (1999). Fracture of the tibial tray after total knee arthroplasty. The Journal of Arthroplasty, Elsevier. 14(1), s. 112-114.en_US
dc.identifier.endpage114en_US
dc.identifier.issn1532-8406
dc.identifier.issue1en_US
dc.identifier.scopusqualityQ1en_US
dc.identifier.startpage112en_US
dc.identifier.urihttps://www.sciencedirect.com/science/article/pii/S0883540399902127
dc.identifier.urihttps://hdl.handle.net/20.500.12415/4375
dc.identifier.volume14en_US
dc.institutionauthorÜğütmen, Ender
dc.language.isoenen_US
dc.publisherElsevieren_US
dc.relation.ispartofThe Journal of Arthroplastyen_US
dc.relation.isversionof10.1016/S0883-5403(99)90212-7en_US
dc.relation.publicationcategoryUluslararası Hakemli Dergide Makale - Kurum Öğretim Elemanıen_US
dc.rightsinfo:eu-repo/semantics/openAccessen_US
dc.snmzKY03536
dc.subjectkneeen_US
dc.subjectarthroplastyen_US
dc.subjectprosthesisen_US
dc.subjectfractureen_US
dc.subjectbreakageen_US
dc.subjectcomplicationen_US
dc.titleFracture of the tibial tray after total knee arthroplastyen_US